YOKOGAWA DIFFERENTIAL PROBE 100 MHZ/1400 V 700924

Category: VOLTAGE PROBES

Frequency bandwidth:  DC to 100 MHz

Maximum continuous input range:  1400 Volts
Attenuation Ratios:  100:1 or 1000:1
 
Functions with Digital Oscilloscopes, ScopeCorders or other waveform measuring instruments.
  • DETAILS
    • FEATURES

      Applicable for all Digital Oscilloscopes or ScopeCorders.

      With increases in the speed of measured signals in recent years, it has become necessary to make floating measurements over a wide frequency band. The 700924 differential probe lets you make wide-band (100 MHz) differential input measurements by just connecting the probe to the input of a single-end input digital oscilloscope.

      Oscilloscope input amplifiers are normally single-ended. They do not provide accurate measurements when used in combination with a 10:1 passive probe for measuring the gate-to-source voltage of a switching power supply, or potential difference with a common mode voltage in elements such as Insulated Gate Bipolar Transistors (IGBTs).

      Furthermore, there is a risk that current will enter through the oscilloscope ground line and damage the oscilloscope.

      Tm 700924 02 1

      Tm 700924 03 1

    • SPECIFICATIONS
       
      Input type Balanced differential input
      Frequency band DC up to 100 MHz (-3 dB)
      Input attenuation ratio Switched between 1/100 and 1/1000
      Input impedance Approximately 4 MΩ, approximately 10 pF parallel
      Gain accuracy ±2% (common mode voltage: 400 V or less)
      ±3% (common mode voltage: 1000 V or less)
      Maximum allowed
      differential voltage
      ±350 V (DC + ACpeak) or 250 Vrms (1/100 range)
      ±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range)
      Maximum common mode
      input voltage
      ±1400 V (DC + ACpeak) or 1000 Vrms (both 1/100 and 1/1000 ranges)
      Maximum input voltage ±1400 V (DC + ACpeak) or 1000 Vrms (1/1000 range)
      Common Mode Rejection Ratio (CMRR) -80 dB (60 Hz), -50 dB (1 MHz)
      Output voltage ±3.5 V
      Output impedance Used with 1 MΩ input impedance
      Power supplies Dry cells, four R6P (SUM-3); when using the AC/DC adapter, use an output voltage of 6 V/200 mA or greater/center plus
      Dry cell life Approximately 2 hours under continuous use
      External dimensions (approximate) 207 mm x 83 mm x 38 mm¹
      Total length Approximately 1.5 meters
      Weight Approximately 800 grams (excluding dry cells)
      Approved standards EMC: EN55011: 1991 + A1: 1997 + A2: 1996
      EN50082-1: 1992
      Safety: EN61010-1: 1993 + A2: 1995
      EN61010-2-031: 1994

      ¹: 38 mm is listed as the maximum value.
      Note: Dry cells and AC/DC adapter not included.

      Model and Code

      Model NameCodeDescription
      Differential probe 700924 DC to 100 MHz band

      Each 700924 differential probe comes standard with red and black pincher tips (one each).

Location